Measurement of all-trans-retinol and 13-cis-retinol
Revision as of 12:29, 8 January 2018 by Admin (talk | contribs) (Protected "Measurement of all-trans-retinol and 13-cis-retinol" ([Edit=Allow only administrators] (indefinite) [Move=Allow only administrators] (indefinite)))
Contents
Golden standard
EN 12823-1:2000 Foodstuffs - Determination of vitamin A by high performance liquid chromatography - Part 1: Measurement of all-trans-retinol and 13-cis-retinol
- draft prEN 12823-1:2012
Method indicator
- Name
- Code
Scope
The European Standard EN 12823-1 describes a method for the determination of vitamin A from foodstuffs by high performance liquid chromatography (HPLC). This part 1 of the EN 12823 specifies a method for the measurement of all-trans-retinol and 13-cis-retinol.
Principle
Sample pretreatments include saponification and extraction of retinol. The measurement is carried out by HPLC using either fluorometric or ultra-violet detection.
Key steps
Saponification and extraction
- The samples should be homogenised prior to saponification and extraction.
- Saponification is performed by adding methanol or ethanol, antioxidant and potassium hydroxide solution to the sample solution.
- The solution is heated to 80-100 °C for 15-45 min or kept in room temperature overnight.
- Extraction is carried out with suitable solvent, and the extracts are washed with water.
- The extract is evaporated and dissolved using mobile phase.
HPLC
- Liquid chromatographic system consisting of a pump, an injector, a UV-VIS (325 nm) or fluorescence detector (excitation wavelength 325 nm, emission wavelength 475 nm) and a data evaluation system.
- Normal phase column: particle size 5 µm, diameter 4 mm, length 125-250 mm.
- Mobile phase: n-hexane:n-butanol, 98:2 (v/v)
- Also other chromatographic systems can be used if equivalent results are guaranteed.
- The performance criteria is the baseline separation of all-trans-retinol and 13-cis-retinol from interferences.
Identification and detection
- All-trans-retinol and 13-cis-retinol are detected either fluorometrically (excitation wavelength 325 nm, emission wavelength 475 nm) or with a UV detector (325 nm).
- Identification of all-trans-retinol and 13-cis-retinol is done by the comparison of the retention time obtained with the standard test solution to that of the sample test solution.
- Identification can also be done by adding the standard substances to the sample test solution.
- Identification can also be done by adding the standard substances to the sample test solution.
Quantification and calculations
- If quantification is done by external standard method, peak areas or heights obtained with sample solutions are compared to the corresponding values of the standard substance.
- Also internal standard method may be used if the similar behaviour of the internal standard and the analyte itself during the analysis has been tested.
- The concentrations of all-trans-retinol and 13-cis-retinol in mg/100 g of the sample are calculated.
Remarks
- After saponification, the extract can be used for the determination of beta-carotene using a method described in EN 12823-2.
- The concentration of calibration solution must be checked spectrometrically.
- The samples should be protected from excessive UV radiation, light, acidity and heat treatments.
- Vitamin A is sensitive especially to UV radiation and oxidising agents.
- The standard solutions must be stored at below 4 °C and protected from light.
- When analysing samples with high starch content, treatment with taka-diastase prior to saponification may be needed.
Criteria for analytical performance and analytical quality control
- Greenfield and Southgate discuss the criteria for analytical performance and quality of analytical data in their book.